可是说明书找不到了,按了上面的键似乎没法自己设程序,只有原先的一些预设程序。
还有,那些溶液是不是可以通用的啊,有没有配方啊。。。
我是新手。。。
有知道的麻烦告诉我下,很感谢哈
洗板机洗板是将手工操作改由洗板机进行,人为因素少,速度快,条件恒定,但是使用洗板机洗板的一个特点是,每次洗板后不能拍干,故有较多的液体残留,需增加洗板次数来减少这种误差。洗板机洗板次数越少,假阳性越高,本实验中,洗板1次,假阳性率高达90%;洗板2次,假阳性率为38%;洗板3次,假阳性率为2%;洗板4~5次,未出现假阳性。洗板次数并非越多越好,太多不但浪费人力、物力,并且也会导致假阴性结果,所以一般试剂盒要求洗板4~5次即可。洗板机洗板时应经常检查冲洗头是否通畅,若被杂物堵塞时可用注射器针头挑出。为了洗涤效果较好,实验室可采用机器与人工洗涤相结合的方法,在机器洗涤后再进行人工洗板1~2次,能达到理想的洗涤效果。
综上所述,在ELISA快速法检验中,手工法洗板和洗板机洗板没有本质的差别,只要操作合乎要求,均能得到正确、可靠的结果。展开