Pointe180impulse™
Catalog#:90.P3900-180
GENERALINFORMATION
ThePointe180impulse™isanISEElectrolyteAnalyzerspecificallydesignedforthedirectdeterminationofNa+,K+andCl-byionselectiveelectrodesinwholeblood,serum,andaqueousQCmaterial.
Reagentpacks,electrodes,andtheinstrumentitselfallcontainuniqueelectronicsallowingforselfcalibrationandmonitoringfortroubleshooting.
Theanalyzerisabottle-free,standalone,benchtop,ISEelectrolyteanalyzerdesignedforthedirectdeterminationofNa+,K+andCl-byionselectiveelectrodes.Theunitcontainsabuiltinthermaldotmatrixprinter.Theunitisprogrammablethroughcustomsoftwareallowingtheusertodefineprogramsfollowingtheirownestablishedprotocols.Thesoftwareisdisplayedonaninteractivetouchscreenwhichguidestheuserstep-by-stepthroughtheprocessusingonscreenprompts,errormessagesandgraphicaldisplays.
Reagentsaresuppliedinpackswhichplugintothesideoftheunitwithnousercontacttowastematerial.Theunitwilltrackreagentexpirationandquantityofreagentremaining.*
*NotavailableforclinicalhumanbloodtestingintheUS
ORDERING INFORMATION
Catalog#: | Product: |
90.P3900-10 | Pointe180impulse™ |
RELATEDPRODUCTS
Catalog#: | Product: |
90.B11-100 | ReagentPack |
90.B11-110 | Na+Electrode |
90.B11-120 | K+Electrode |
90.B11-130 | Cl-Electrode |
90.B11-140 | ReferenceElectrode |
90.B11-310 | Tri-LevelQCKit |
FEATURES
Simplicity
- Stepbystepuserprompts
- InteractivetouchscreenLCDwithcolorsandgraphics
- Self-calibratingandmonitoring
- Resultsmaybeprinted,storedon-board,ordownloadedtoathumbdrive
- DownloadabletroubleshootingABIlityforoff-sitesupportandanalysis
IntelligentUserInterface
- Pointe180impulsecanbeeasilyprogrammedtoconformtoestablishedlaboratoryprotocols
- Thesoftwareallowsselectionofavarietyofperformanceoptions
- Patientandqualitycontrolreferencerangesandpatientandoperatoridentificationnumbersareconvenientlyenteredviathenumerickeypad
- Intuitiveuser-friendlyscreenprompts
- AutomaticallystoresdataforLevey-Jenningsplots
ConvenientandSafe
- "Smart"softwareprovideswarnings,reminders,anderrormessages
- Environmentallyfriendlymaterials
- Self-containedwastebagwithsafetyvalveprotection
- Detailedusermanual
- Fullytestedforelectricalsafety
SPECIFICATIONS
Method: | Directmeasurementbyionselectiveelectrode |
SampleType: | WholeBlood,Serum,AqueousQCMaterial |
SampleSize: | 60µl±5µl |
SampleRate: | 48/hourmaximumwithoutprinting |
| 45/hourmaximumwithprinting |
MeasurementRanges: | Sodium — 40-205mmol/L |
| Potassium — 1.5-15.0mmol/L |
| Chloride — 50-200mmol/L |
| |
SampleProbe: | 316StainlessSteel |
Vessels: | Capillarytubeswithadaptor,tubesupto100mmdepth,syringes,samplecups |
Electrodes: | Sodium(Na+)Electrode |
| Potassium(K+)Electrode |
| Chloride(Cl-)Electrode |
AnalysisTime: | Typically<60seconds |
PatientDataStorage: | 200results(withcapabilitytoassigna15characterpatientidentifier) |
QCDataStorage: | 270reADIngs(30readingseachof3controlsforeachofthe3electrodes,includesLevey-Jennings) |
Calibration: | Ondemandorautomaticfrom1to8hourintervals |
PowerRequirements: | 100-240VAC,50-60Hz90Wmaximum |
Dimensions: | 14”wx12”hx8”d(35.5cmx30.5cmx20cm) |
Weight(withreagentpack): | 14.5lbs(6.6kg) |
Display: | Interactivetouchscreen3.5”(9cm)LCDcolorgraphicdisplay |
Interface: | IndirectviaUSBDrive(Included) |
Printer: | Thermaldotmatrixwithgraphiccapability,29characterwidth |
ebiomall.com






>
>
>
>
>
>
>
>
>
>
>
>